When measuring surface tension using the SensDyne maximum differential bubble
pressure method, as the viscosity of the fluid increases, the hydrodynamic resistance
against the formation of the bubble also increases. This causes a measurement error
referred to as Viscosity Effect. Though negligible for coatings of small viscosity, it can
be several tens of dynes/cm for highly viscous coatings.
For non-viscous fluids the correction does not exceed normal instrument measurement
error of +/- 0.1 Dynes/cm. For very viscous liquids, however, the value can reach up to
tens of Dynes/cm at small surface age values, and correction factors are needed to obtain
true viscosity compensated values for dynamic surface tension, unless an automatic
viscosity compensating method is used.
For all Sensadyne Tensiometers that have two or more probes there is a method to
eliminate the Viscosity Effect. For manual two probe tensiometers, such as the QC-series,
the Windows® software (version 1.4.1) has an "Individual Data Recording" option as
described in this excerpt from the Software Manual:
"Individual Data Recording: This option should only be used when the tensiometer is
used to measure very viscous fluids. This option collects data from each individual, valid
detected peak when the tensiometer is set up for the viscosity compensation mode using
Stokes Law. [The large orifice is set to a surface age equal to the ratio of the large to
small orifice - for a 4.0 mm. and 1.0 mm. orifice the surface age of the large orifice is set
four times faster than the surface age of the small orifice]. Minimum values will occur
when both orifices release bubbles at the same time and the viscosity effect at each
orifice cancels out. The lowest readings collected in the data file set (after running
perhaps forty or fifty scans) will reflect the true viscosity compensated surface tension."
The SensaDyne PC500-LV uses three probes, one 0.5mm, one 1.0mm, and one 4.0mm.
The 4.0mm acts as a large control probe shared with both small orifices. This allows
simultaneous but separate differential pressure signals to be generated by each of the
smaller orifices. The maximum bubble pressure peak values from each of these
differential signals are electronically substracted by the software to give a final
differential in which the hydrodynamic effect cancels. This patented process makes it
simple for the user to obtain a viscosity compensated value without the use of correction
factors.
Copyright © 1981-2023 SensaDyne Instrument Div. All rights reserved.